Chapter 8 ____________________________________________________________ Technical Data
VAISALA ______________________________________________________________________ 169
Dewpoint Temperature (HMT337 Warmed
Probe Option)
Find the intersection of the dewpoint temperature curve and the dewpoint
difference reading (process temperature-dewpoint temperature) on the x-
axis and read the accuracy in dewpoint measurement on the y-axis.
0508-017
Figure 81 Accuracy in Dewpoint Measurement
Operating Environment
Operating temperature range
for humidity measurement -70 ... +180 °C (-94 ... +356 °F)
see probe specifications
for transmitter body electronics -40 ... +60 °C (40 ... +140 °F)
with display 0 ... +60 °C (+32 ... +140 °F)
Storage temperature range
without display -55 … +80 °C (-67 … +176 °F)
with display -40 … +80 °C (-40 … +176 °F)
Electromagnetic compatibility EN61326-1: Electrical equipment for
measurement, control, and laboratory use –
EMC requirements – for use in industrial
locations