User's Guide _______________________________________________________________________
62 ___________________________________________________________________ M211060EN-F
Changing the HUMICAP® 180R Sensor
(HMP110, HMP110D and HMP113)
NOTE
Replacing the humidity sensor of the HMP110
, HMP110D and HMP113
is not necessary in normal operation. If the accuracy of the probe does
not seem to be within specification, it is likely that the accuracy can be
restored by performing
the adjustment procedure. See section Two-Point
Adjustment of RH Measurement (HMP110, HMP110D and HMP113)
on page
56.
Follow this procedure to replace the humidity sensor of the HMP110,
HMP110D and HMP113 in case it has been damaged, or normal
adjustment is not sufficient to restore the measurement accuracy.
Calibration and adjustment of the humidity measurement is required after
the sensor change.
To perform this procedure, you need a new HUMICAP® 180R sensor. It
is also recommended that you replace the filter with a new one. For
order codes, see section Options and Accessories on page 70.
1. Remove the filter from the probe by turning it counter-clockwise.
2. There are two sensors under the filter, the HUMICAP® sensor and
a temperature sensor. Identify the HUMICAP® sensor – do not
touch the temperature sensor.
1210-035
Figure 26 HUMICAP® 180R Sensor
3. Pull out the old HUMICAP® 180R sensor and insert a new one.
Refer to Figure 25 on page 61. Handle the new sensor by the plastic
frame. DO NOT TOUCH THE SENSOR PLATE.
4. Perform a two-point adjustment of the RH measurement as
instructed in section Two-Point Adjustment of RH Measurement
(HMP110, HMP110D and HMP113) on page 56.
5. Attach a new filter on the probe.