Vaisala DRYCAP DMT340 SERIES Podręcznik Użytkownika Strona 49

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 169
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 48
Chapter 3 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 49
Leak Screw Installation
When the probe cannot be installed directly in the pressurized process or
process pipe, a leak screw installation can be used.
In this installation method, you must install the DMT348 probe using a
fitting body with leak screw, see Figure 27 on page 44. A small sample
flow from the process goes through the probe and via the leak screw out
to atmospheric pressure enabling a fast response time although the probe
is not installed in the process.
0503-036
Figure 31 Probe in Leak Screw Installation
Numbers refer to Figure 31 above:
1 = Probe
2 = Filter
3 = Ball of the ball valve
4 = Leak screw
Przeglądanie stron 48