USER'S GUIDE_______________________________________________________________________
18 ___________________________________________________________________ M210195EN-A
INTV Setting the output interval
INTV x y <cr>
where:
x = output interval (0...255)
y = unit (s, min, h)
This command is used to set and inspect the output interval. The R
command is used to start the outputting.
Examples:
>intv <cr>
Output intrv. : 0 s
>
>intv 10 s <cr>
Output intrv. : 10 s
>r <cr>
NOTE
In case of the half duplex RS 485/422 serial communication, the user
is requested to set the interval time to one second or more. This
enables an interruption of outputting, if the R command is given by
mistake.
ADDR Setting the transmitter address (for POLL mode)
ADDR x <cr>
where:
x = the address (0 ... 99)
This command is used to set and inspect the address of the transmitter
for the POLL mode. The address feature is important when multiple
transmitters are connected to one RS232C interface or when a
transmitter is used on an RS485/422 serial interface.
Example of setting the address to 7:
>addr 7 <cr>
Address : 7
>