Vaisala INTERCAP HMDW80 SERIES Dokumentacja Strona 82

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 187
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 81
80
/ TECHNOLOGY DESCRIPTION
Vaisala DRYCAP® Sensor for Measuring
Dew Point
DRYCAP in Brief
Thin-film polymer sensor with
unique auto-calibration function
Wide measurement range, dew
point measurement down to
-80 °C (-112 °F)
Accuracy ± 2 °C (± 3.6 °F)
NIST-traceable dew point
measurement
DRYCAP’s Unique Benefits
Excellent long-term stability,
with recommended 2-year
calibration interval
Rapid response time
Withstands condensation and
recovers rapidly
Resistant to particulate
contamination, oil vapor, and
most chemicals
In 1997 Vaisala introduced DRYCAP
®
, a new type of dew point sensor based on thin-film polymer
technology. Since its launch, the DRYCAP product family has grown to encompass a huge range
of applications, from drying processes to compressed air and dry chambers. The DRYCAP sensor
is particularly renowned for its reliable performance in hot and very dry environments.
How It Works
DRYCAP’s unrivalled performance is
based on two innovations: the proven
capacitive thin-film polymer sensor
and the auto-calibration function.
The sensor’s thin-film polymer
absorbs or releases water vapor as
the surrounding humidity increases
or decreases. The dielectric
properties of the polymer change
as the humidity around the sensor
changes, as does the capacitance of
the sensor. Capacitance is converted
into a humidity reading. The
capacitive polymer sensor is bonded
together with a temperature sensor,
and dew point is calculated from the
humidity and temperature readings.
Vaisalas patented auto-calibration
function optimizes the measurement
stability at low dew points.
The sensor is heated at regular
intervals during the automated
auto-calibration procedure. The
humidity and temperature readings
are monitored as the sensor cools
to ambient temperature, with offset
correction compensating for any
potential drift. This enables the
DRYCAP sensor to deliver accurate
measurements in the long term,
dramatically reducing the need for
maintenance.
Typical Applications for
Dew Point Measurement
Vaisala DRYCAP dew point
instruments measure dew point in
industrial applications, where gas
humidity is typically very low. Dew
point is often a critical parameter,
with inadequate control resulting in
problems such as process downtime,
damaged process equipment, and
deterioration in end-product quality.
Dew point is measured in various
drying and heat-treatment processes.
It is also controlled in compressed
air, where excess moisture can result
in poor end-product quality, ice
formation, and equipment corrosion.
Dew Point
Przeglądanie stron 81
1 2 ... 77 78 79 80 81 82 83 84 85 86 87 ... 186 187

Komentarze do niniejszej Instrukcji

Brak uwag