Technical Data
Weather
Please contact us at
www.vaisala.com/requestinfo
www.vaisala.com
Scan the code for
more information
Barometric Pressure
Range 600 ... 1100 hPa
Accuracy ±0.5 hPa at 0 ... +30 °C (+32 ... +86 °F)
±1 hPa at -52 ... +60 °C (-60 ... +140 °F)
Output resolutions and units 0.1 hPa, 10 Pa, 0.0001 bar,
0.1 mmHg, 0.01 inHg
Relative Humidity
Range 0 ... 100 %RH
Accuracy ±3 %RH within 0 ... 90 %RH
±5 %RH within 90 ... 100 %RH
Output resolution and unit 0.1 %RH
General
Operating temperature -52 ... +60 °C (-60 ... +140 °F)
Storage temperature -60 ... +70 °C (-76 ... +158 °F)
Operating voltage 5 ... 32 VDC
Typical power consumption 3 mA at 12 VDC (with defaults)
Heating voltage 5 ... 32 VDC / 5 ... 30 VAC
RMS
)
Serial data interface SDI-12, RS-232, RS-485, RS-422,
USB connection,
Weight 650 g (1.43 lb)
Housing IP65
Housing with mounting kit IP66
Electromagnetic Compatibility
Complies with EMC standard EN61326-1; Industrial Environment
IEC standards IEC 60945/61000-4-2 ... 61000-4-6
Liquid Precipitation
RAINFALL
cumulative accumulation after the
latest automatic or manual reset
output resolutions and units 0.01 mm, 0.001 inches
accuracy 5%*
RAINFALL DURATION
counting each ten-second increment
whenever water droplet is detected
output resolution and unit 10 s
RAIN INTENSITY
one-minute running average in
ten-second steps
range 0 ... 200 mm/h (broader range with
reduced accuracy)
output resolutions and units 0.1 mm/h, 0.01 inches/h
HAIL
cumulative amount of hits against
the collecting surface
output resolutions and units 0.1 hits/cm
2
, 0.01 hits/in
2
, 1 hits
HAIL DURATION
counting each ten-second increment
whenever hailstone is detected
output resolution and unit 10 s
HAIL INTENSITY
one-minute running average
in ten-second steps
output resolutions and units 0.1 hits/cm
2
h, 1 hits/in
2
h, 1 hits/h
* Due to the nature of the phenomenon, deviations caused by spatial
variations may exist in precipitation readings, especially in a short time scale.
The accuracy specification does not include possible wind induced errors.
Δ
°C
°C
-0.7
0.7
0.6
0.5
0.4
0.3
0.2
0.1
-0.1
-0.2
-0.3
-0.4
-0.5
-0.6
0
0-60
-40
-20 20
40
60
80
-80
Air Temperature
Range -52 ... +60 °C (-60 ... +140 °F)
Accuracy for sensor at +20 °C ±0.3 °C (±0.5 °F)
Accuracy over temperature range (see graph below)
Output resolutions and units 0.1 °C, 0.1 °F
Ref. B210417EN-K ©Vaisala 2014
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Wind
SPEED
range 0 ... 60 m/s
response time 250 ms
accuracy ±3% at 10m/s
output resolutions and 0.1 m/s, 0.1km/h,
units 0.1 mph, 0.1 knots
DIRECTION
azimuth 0 ... 360°
response time 250 ms
accuracy ±3°
output resolution and unit