User's Guide _______________________________________________________________________
26 ___________________________________________________________________ M210474EN-C
MMT317 Probe with Swagelok Connector for Tight-
place Installations
The Swagelok installation kit for MMT317 includes a Swagelok
connector with ISO3/8" thread (Vaisala order code: SWG12ISO38) or
NPT1/2" thread (Vaisala order code: SWG12NPT12).
0509-144
Figure 10 MMT317 Probe with Swagelok Installation Kit
0509-150
Figure 11 MMT317 Probe Installation to Pipeline with Swagelok
Installation Kit
The following numbers refer to Figure 11 above:
1
=
Probe
2
=
Duct connector
3
=
ISO3/8" or NPT1/2" thread
4
=
Swagelok connector
5
=
Ferrules
6
=
Upper edge of the connector nut
7
=
Upper edge of the probe