Chapter 3 ________________________________________________________________ Installation
VAISALA ________________________________________________________________________ 21
The following two fitting body options are available:
- Fitting Body ISO1/2 solid structure (DRW212076SP)
- Fitting Body NPT1/2 solid structure (NPTFITBODASP)
Table 4 MMT318 Pipe Dimensions
Use teflon tape or thread sealant to seal the connection between the
fitting body and the process/ball valve.
0507-025
Figure 6 Sealing of Fitting Body into Process
Tightening the Clasp Nut
1. Adjust the probe to a suitable depth according to the type of
installation.
2. Tighten the clasp nut first manually.
3. Mark the fitting screw and the clasp nut.
4. Tighten the nut a further 50 - 60º (ca. 1/6 turn) with a wrench. If
you have suitable torque wrench, tighten the nut to max 45 ± 5 Nm
(33 ± 4 ft-lbs).