Ref. B210769EN-C ©Vaisala 2009
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifi cations — technical included — are subject
to change without notice.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Technical data
Electrical connections
Two analog outputs,
selectable and scalable
0 ... 20 mA or 4 ... 20 mA
Typical accuracy of analog output at +20 °C
+0.05 % full scale
Typical temperature dependence
of analog output
0.005 %/°C (0.003 %/°F)
of full scale
Serial output RS--232C
Connections M12 8-pole connector with RS-232C,
current outputs (two channels) and U
in
Operating voltage 12 ... 35 VDC, the maximum
operating voltage for a device with
sensor heating is 24 VDC
Power consumption 30 mA with RS-232
External load R
l
< 500 Ohm
Startup time after powerup 3 s
General
Operating temperature range for
electronics -40 ... +60 °C (-40 ... +140 °F)
Storage temperature range -55 .... +80 °C (-67 ... +176 °C)
Operating pressure
HMT314 0 ... 100 bar
HMT318 0 ... 40 bar
HMT315, HMT317 vapor tight
Transmitter housing material G-AlSi10Mg
Transmitter base material ABS/PC
Housing classifi cation IP65
Cable feed through
alternatives
8-pole connector with 5 m cable,
Female 8-pin connector screw joint for
cable diameter 4 ... 8 mm
Sensor protection PPS grid with stainless steel net,
PPS grid, Sintered fi lter
Membrane stainless steel fi lter
Complies with EMC standard EN61326-1, Industrial environment
Note: When using the current output, the RF fi eld susceptibility
level according to standard EN61000-4-3 with a frequency band
of 110 ... 165 MHz, is only 3V/m (generic environment) with the
specifi ed accuracy.
Measured values
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Sensor
Vaisala HUMICAP
®
180R typical applications
Vaisala HUMICAP
®
180RC applications with chemical
purge and/or warmed probe
Accuracy (incl. non-linearity, hysteresis and repeatability)
at a temperature range of
+15 ... +25 °C ±1 %RH (0 ... 90 %RH)
±1.7 %RH (90 ... 100 %RH)
-20 ... +40 °C ±(1.0 + 0.008 x reading) %RH
-40 ... +180 °C ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty
(+20 °C)
±0.6 %RH (0 ... 40 %RH)*
±1.0 %RH (40 ... 97 %RH)*
* Defi ned as ±2 standard deviation limits. Small variations possible, see
also calibration certificate.
Response time (90 %) at +20 °C 17 s with grid fi lter
in 0.1 m/s air fl ow 50 s with grid and steel, netting fi lter
60 s with sintered fi lter
TEMPERATURE
HMT311 -40 ... +60 °C (-40 ... +140 °F)
HMT313 -40 ... +80 °C (-40 ... +176 °F)
or -40 ... +120 °C (-40 ... +248 °F)
HMT314, HMT315, HMT317, HMT318 -70 ... +180 °C (-94 ... +356 °F)
Accuracy over temperature range (see graph below)
Typical temperature
dependence of electronics
+0.05 °C/°C (+0.005 °F/°F)
Temperature sensor Pt100 IEC751/3 class B
6
°C
°C
-0.7
0.7
0.6
0.5
0.4
0.3
0.2
0.1
-0.1
-0.2
-0.3
-0.4
-0.5
-0.6
0
0-60 -40 -20 20 40 60 80 100
120
140
160
180
-80
HUMICAP
®
is a registered trademark of Vaisala.