12
°C
°C
0
0-80 -60 -40 -20 20 40 60 80 100 120 140 160
180
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0.1
0.2
0.3
0.4
0.5
0.6
0.7
Temperature sensor PT 100 RTD 1/3 Class B IEC 751
Other variables available (depends on model)
dewpoint temperature, mixing ratio, absolute humidity, wet bulb
temperature, enthalpy, water vapor pressure
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Accuracy (including non-linearity, hysteresis, and repeatability)
with Vaisala HUMICAP
®
180 or 180R for typical applications
with Vaisala HUMICAP
®
180C or 180RC for applications with
chemical purge/warmed probe
at +15 ... +25 °C (59 ... +77 °F) ±1 %RH (0 ... 90 %)
±1.7 %RH (90 ... 100 %RH)
at -20 ... +40 °C (-4 ... +104 °F) ±(1.0 + 0.008 x reading) %RH
at -40 ... +180 °C (-40 ... +356 °F) ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C) ± 0.6 %RH (0 ... 40 %RH)
± 1.0 %RH (40 ... 97 %RH)
(Defi ned as ±2 standard deviation limits. Small
variations possible, see also calibration certifi cate.)
Response time (90 %) at +20 °C (+68 °F) 8 s / 17 s* with grid fi lter
in still air 20 s / 50 s* with grid + steel netting fi lter
40 s / 60 s* with sintered fi lter
* with HUMICAP
®
180R or 180RC sensor
TEMPERATURE
Accuracy at +20 °C (+68 °F) ± 0.2 °C (± 0.36 °F)
Accuracy over temperature range (measurement range
depends on model)
Inputs and outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC, 50/60 HZ
Power consumption at +20 °C (U
in
24 VDC)
RS-232 max. 25 mA
U
out
2 x 0 ... 1 V/0 ... 5 V/0 ... 10 V max. 25 mA
I
out
2 x 0 ... 20 mA max. 60 mA
display and backlight + 20 mA
during chemical purge max. 110 mA
during probe heating (HMT337) + 120 mA
Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0... 1 V, 0... 5 V, 0... 10 V
Accuracy of analog outputs at +20 °C ±0.05% full scale
Temperature dependence of the
analog outputs ±0.005%/°C full scale
External loads
current outputs R
L
< 500 ohm
0 ... 1 V output R
L
> 2 kohm
0 ... 5 V and 0 ... 10 V outputs R
L
> 10 kohm
Max. wire size 0.5 mm
2
(AWG 20)
stranded wires recommended
Digital outputs RS-232, RS-485 (optional)
Service connection RS-232, USB
Relay outputs (optional) 0.5 A, 250 VAC
Technical data
Ethernet interface (optional)
Supported standards 10/100Base-T
Connector RJ45
Protocols Telnet
Software support Vaisala MI70 link
WLAN interface (optional)
Supported standards 802.11b
Antenna connector type RP-SMA
Protocols Telnet
Security WEP 64/128,WPA
Software support Vaisala MI70 link
Authentication / Encryption (WLAN)
Open / no encryption
Open / WEP
WPA Pre shared key / TKIP
WPA Pre shared key / CCMP (a.k.a. WPA2)
Optional data logger with real-time clock
Logged parameters max. three with trend/min/max values
Logging interval 10 sec (fi xed)
Max. logging period 4 years 5 months
Logged points 13.7 million points per parameter
Battery lifetime min. 5 years
Display LCD with backlight, graphic trend display
of any parameter
Menu languages English, Chinese, Japanese, Spanish, German,
French, Swedish, Russian, Finnish
Komentarze do niniejszej Instrukcji