Vaisala HMP228 Podręcznik Użytkownika Strona 20

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 71
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 19
USER'S GUIDE____________________________________________________________________
3. Make sure that the temperature at the measurement point is
equal to that of the process, otherwise the moisture reading may
be incorrect.
0605-085
Figure 7 Ball Valve Installation
The following numbes refer to Figure 7 on page 18 above.
1 = Probe
2 = Clasp nut. The probe slides easily in place when you first fix
the clasp nut manually. Then complete the installation and
make it stable by tightening the clasp nut with a plastic
wrench. Do not overtighten the screw.
3 = Handle of the ball valve
4 = Process chamber/pipeline
5 = Manual press tool
6 = The groove on the probe indicates the upper adjustment limit
7 = Ball of the ball valve
18 __________________________________________________________________ M210498EN-C
Przeglądanie stron 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 70 71

Komentarze do niniejszej Instrukcji

Brak uwag