CHAPTER 6__________________________________________________________ SERIAL COMMANDS
VAISALA _________________________________________________________________________ 17
setting is 1013 hPa. The ambient pressure value can be set by using
the following command:
MF_PRESSURE xxxx<cr>
where: xxxx = pressure in hPa
Example of setting desired pressure:
>MF_PRESSURE 900
MF_PRESSURE=900
>
Save the settings:
MAIN_SAVE F<cr>
Altitude/Atmospheric pressure table
Altitude m (ft) Atmospheric pressure
(hPa)
Atmospheric pressure
(psi)
0 (sea level) 1013 14.69
500 m (1640 ft) 954 13.84
1000 m (3281 ft) 899 13.04
1500 m (4921 ft) 845 12.26
2000 m (6562 ft) 795 11.53
2500 m (8202 ft) 747 10.83
3000 m (9843 ft) 701 10.17
Temperature setting
The ambient temperature value can be set by using the following
command:
MF_TEMP xxx<cr>
where: xxx = ambient temperature (°C) x 10.
Example of setting desired temperature (25 °C):
>MF_TEMP 250
MF_TEMP=250
>
Save the settings:
MAIN_SAVE F<cr>