USER'S GUIDE____________________________________________________________________
26 __________________________________________________________________ M210855EN-D
Grounding the Cables
To ensure the best possible EMC performance, you must ground the screen of the electrical cable as shown in
1212-005
Figure 16 below and in the instructions below that.
1212-005
Figure 16 Grounding the Screen of Electrical Cable
To ground the electrical cables:
1. Cut back outer sheath to desired length.
2. Cut back screen braiding or screen foil to dimension X (see
1212-005
Figure 16 above).
3. Push the domed cap nut (item 1) and the seal insert with contact
socket of the gland (item 2+3) onto the cable as shown in the
diagram.
4. Bend over the screen braiding or screen foil by about 90° (item
4).
5. Push the seal insert with the contact socket of the gland (item
2+3) up to the screen braiding or screen foil.