USER'S GUIDE_______________________________________________________________________
38 ___________________________________________________________________ M210195EN-A
Humidity
Calibration of the HMP45A-P/D probes should be performed at
regular intervals, depending on the conditions of use and desired
accuracy. The recommended calibration interval is one year.
The HMP45A-P/D probes are easy to maintain, calibrate and adjust.
The probe consists of a probe head and a handle with cable. All
calibration electronics are in the probe head, which can be
disconnected from the handle without disconnecting the wires. The
probe heads are interchangeable.
D
W
T
pull the probe head
off the handle
adjustment trimmers:
W = wet D = dry
(T for factory use only!)
FIGURE 6-2 Adjustment trimmers and probe head connection/
disconnection
Humidity adjustment
For a high-accuracy two-point adjustment, use a Vaisala HMK15 or
HMK13B calibrator and saturated salt solutions as described in the
respective manuals. If there is no local display, read the outputs via
the serial line.
Leave the calibrator and the probe head in the same space for at least
four hours so that their temperatures have time to equalize. Unscrew
the plastic grid of the probe.
The adjustment is done first for the dry end and then for the wet end
with trimmer potentiometers marked “D” (dry, <50 %RH) and “W”
(wet, >50 %RH). The potentiometers are located under a protective
plug; see FIGURE 6-2. Use a ceramic screwdriver with 2.5 mm blade for
adjusting the potentiometers. Note: if zero point is adjusted in
Nitrogen (N
2
), the minimum output signal of 0.008 V corresponds to a
relative humidity of 0.8 %RH.
TABLE 6-2 Greenspan’s calibration table
Temperature °C 15 20 25 30 35
LiCl %RH *) 11.3 11.3 11.3 11.3
NaCl %RH 75.6 75.5 75.3 75.1 74.9
K
2
SO
4
%RH 97.9 97.6 97.3 97.0 96.7
*) Do not use or store the LiCl solution in temperatures below +18°C as its humidity
equilibrium may change permanently
Przeglądanie stron 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 63 64

Komentarze do niniejszej Instrukcji

Brak uwag

Bosch DHL755BUC Handbücher

Bedienungsanleitungen und Benutzerhandbücher für Fans Bosch DHL755BUC.
Wir stellen 3 PDF-Handbücher Bosch DHL755BUC zum kostenlosen herunterladen nach Dokumenttypen zur Verfügung Bedienungsanleitung


Bosch DHL755BUC Bedienungsanleitung (2 Seiten)


Marke: Bosch | Kategorie: Fans | Größe: 0.24 MB |

 

Inhaltsverzeichnis





Weitere Produkte und Handbücher für Fans Bosch

Modelle Dokumententyp
600 CFM Internal Blower DHG601DUC Bedienungsanleitung   Bosch 600 CFM Internal Blower DHG601DUC User Manual, 2 Seiten
600 CFM DHG6015DUC Bedienungsanleitung   Bosch 600 CFM DHG6015DUC User Manual, 2 Seiten
DUH36252UC Bedienungsanleitung   Bosch DUH36252UC User Manual, 2 Seiten
600 CFM DHG6023UC Bedienungsanleitung   Bosch 600 CFM DHG6023UC User Manual, 2 Seiten