Chapter 7 ______________________________________ Analog Output Connection
VAISALA __________________________________________________________37
CHAPTER 7
ANALOG OUTPUT CONNECTION
This chapter describes the selecting and scaling of the analog output
connection.
Selecting and Scaling the Analog Output
0505-233
Figure 13 Selecting Analog Output
To get analog measurement data you need an analog output signal cable,
see section Accessories on page 65. One scalable voltage signal channel
0...1.0V is available. You can scale the output to any output range but it
is recommended to have the scaling within the probe's measuring range
to get accurate measurements.