Vaisala HUMICAP HMP155 Podręcznik Użytkownika Strona 34

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 84
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 33
USER'S GUIDE____________________________________________________________________
32 _______________________________________________________________ M210912EN-C
The settings are temporary, and last only for a single session at a time.
On the next connection HMP155 reverts to the settings last set with the
SERI command.
SMODE
Use the SMODE command to set the user port to permanent STOP,
RUN, POLL or SEND mode.
SMODE [xxxx]
where
xxxx= STOP, RUN, POLL or SEND
Example:
>smode
Serial mode : STOP ? POLL
>
Table 12 Selection of Output Modes
Mode Output Available Commands
STOP Only with the SEND command All (default mode)
RUN Automatic output Only command S
POLL Only with the SEND [addr] command ??, SEND [addr], OPEN [addr]
Selected output mode will be activated after power outages.
INTV
Use the INTV command to set the outputting interval for the RUN mode.
The default value for this command is 2 seconds.
INTV [xxx yyy]
where
xxx = Output interval (0...255) where 0 produces an output whenever a
new measurement is made. Interval 1 s produces repeats of the
previous measurement if no newer measurement is available.
yyy = Unit (s, min or h)
Example:
>intv 1 min
Interval : 1 min
>
Przeglądanie stron 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 83 84

Komentarze do niniejszej Instrukcji

Brak uwag