User's Guide _______________________________________________________________________
48 ____________________________________________________________________M210498en-A
CHAPTER 14
MAINTENANCE
Changing the filter and sensor
1. Screw out the filter from the probe head (figure on page 7).
2. Remove the damaged sensor by pulling it out. Handle the sensor by the plastic
socket.
3. Insert a new sensor.
4. Make a humidity calibration.
5. Screw a new filter on the probe head, take care to tighten the filter properly
(5.1 Nm).
Clean the sensor before storing the MMP78 probe
and before calibration
For cleaning of the probe you need instrument air and heptane liquid. Dry
with instrument air to prevent oxidation of the sensor. The oxidation of the
sensor can cause extended response times or drifting.
1. Blow the probe head (with filter) with instrument air to remove the
remains of the oil.
2. Immerse the probe head into heptane liquid and rinse out the oil.
3. Dry the probe head with instrument air. In case you are going to
calibrate the probe, remove the filter and dry the sensor with instrument
air. Check that the sensor looks clean.
Plastic
socket