Vaisala DRYCAP DMT345 Podręcznik Użytkownika Strona 40

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 157
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 39
USER'S GUIDE____________________________________________________________________
40 __________________________________________________________________ M210762EN-D
Cable Grounding
Ground the screen of the electrical cable properly to achieve the best
possible EMC performance.
0605-027
Figure 25 Grounding the Screen of Electrical Cable
1. Cut back outer sheath to desired length.
2. Cut back screen braiding or screen foil to dimension X (see Figure
25 A above).
3. Push the domed cap nut (item 1 in Figure 25) and the seal insert
with contact socket of the gland (items 2 & 3) onto the cable as
shown in the diagram.
4. Bend over the screen braiding or screen foil by about 90º (item 4).
5. Push the seal insert with the contact socket of the gland
(items 2 & 3) up to the screen braiding or screen foil.
Przeglądanie stron 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 156 157

Komentarze do niniejszej Instrukcji

Brak uwag