Chapter 5 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 49
Set Measured Gas Type
Use the GAS command to set the type of gas that is measured.
GAS [type]<cr>
where
type
=
Measured gas type. Options are:
0 = air
1 = SF
6
Example (change currently measured gas from air to SF
6
):
gas
Gas : 0 ? 1
Set Measurement Filtering
Use the FILT command to view or set the speed at which the latest
measurement result is integrated into the output readings. The command
affects both analog output and serial line output.
FILT [a.aaa]<cr>
where
a.aaa
=
Range 0.001 ... 1.0.
1.0 = No filtering, latest
measurement is output
without averaging
0.5 = Average of last two measurements
0.1 = Average of ca. 16 measurements (default)
Example (show current setting):
filt
Filter : 0.100 ?
Example (set filtering to 0.5):
filt 0.5
Filter : 0.500